Physical Dimensions |
Height |
1306 mm, 51.4 in |
Width |
1673 mm, 46.6 in |
Depth |
1185 mm, 65.8 |
Weight (TBC) |
390kg, 860lb |
System Specifications |
Access loading |
Continuous sample, reagent and biochip input and output |
Aspiration Capability |
Precision pump with liquid level and clot detection, 7-300µl aspiration and dispense range |
Biochip Capacity |
200 Biochips yielding up to 9800 tests containing up to 20 different panels onboard simultaneously |
Biochip Format |
Cartridge format with 10 x single biochip, individually sealed wells |
Calibration Principal |
6 Point Master Curve Generation, individual point re-run |
Connectivity |
6 x USB ports, 1 x Ethernet, 1 x RS232, |
Data Backup Methods |
Via writable DVD, CD, USB Mass-storage or Network Folder |
Data Storage Capacity |
Up to 12 million test results with complete raw data. RAID I data mirroring |
Environmental |
Ambient Environment 15 - 32°C, <2°C temperature variation per hour, 20 – 85% relative humidity |
Liquid Waste |
High and low concentration waste separation |
LIS Connectivity |
Bi-directional: ASTM Standard (RS232 Connection) |
Maximum Throughput |
2940 results per hour |
Measurement Principal |
Chemiluminesence |
Network Services |
Highly secure Remote Diagnostics, automated software update and array updates |
Operational Modes |
Random and continuous access, priority STAT loading, deadline sample completion, group tests by patient, batch mode |
Operator Interface |
22” Touch-screen monitor, highly intuitive e-touch software |
Peripherals |
Handheld, wireless 1D\2D Barcode scanner, A4 Laser Printer |
Quality Control |
Interactive Levey-Jennings, user definable multipoint rules with system alarm |
Reagent Capacity |
20 twin vessel reagent bottles (50ml per vessel) with Butterfly Valve System (BVS) |
Sample Bar-code acceptable Symbologies |
Code 39, Codabar, Code 128, I 2 of 5, Code 93, UPC/EAN, Pharmacode, PDF417 (option), MicroPDF(option) |
Sample Capacity |
100 Samples |
Sample Dilution |
Automatic, on-board dilution, continuously variable between 1:1 and 1:100, array dependent |
Sample Types |
Array dependent, BALF, cell culture supernatant, CSF, feed, forensic matrices, honey, milk, plasma, saliva, serum, tissue, urine |
Sample Tube Compatibility |
2ml Paediatric Cups, Standard Sample Tubes (D. 10-16mm x H. 75-100mm), Cobus Cups (using optional adaptors) |
Sample Racks |
10 Independent racks holding 10 samples each. Each rack can hold any combination of sample tubes |
Sample Volume |
Panel dependent, from 70pl per test |
STAT Capability |
Continuous Access, no dedicated loading route required |
System Fluids |
Onboard dilution of bulk reagents, refrigerated storage, continuous level monitoring |
Time to First Result |
Array dependant, approximately 35 minutes |
Walk-away time |
1 hour 40 minutes |
Power Requirements |
Input Voltage (TBC) |
100-240 VAC at 50-60 Hz, 1500 Watt approx. |
UPS |
Optional |
Water Requirements |
Direct Plumbing |
Maximum pressure of 50 kPa |
Supply |
Direct Plumbing or Internal 20L reservoir tank |
Water Quality |
CLSI Type II or better |