Physical Dimensions
Height
1150 mm/45.28 inches
Depth
800 mm/31.5 inches
Length
1260 mm/49.6 inches
Weight
400kg/881.85 lbs
Performance Characteristics
Analyser Type
Fully automated random access clinical analyser with STAT priority
Throughput
800 tests/hour or up to 1200 tests/hour with ISE
Assay Types
End-point, kinetic, fixed rate, bichromatic, turbidimetric and ISE
Test Channels
44 photometric tests and 3 ISEs can be performed at the same time
Sample Unit
Sample Capacity
Double sample carousel with 140 sample positions including 80 routine sample positions, 10 STAT, 39 calibrator, 8 QC and 3 wash positions
Sample Cooling
Continuous cooling of QC carousel 5-15oC
Sample Tubes
Supports various size sample tubes; specification: (12 - 16mm) x (25 - 100mm)
Sample Identification
Barcode sample identification code 128, code 39
Sample Volume
2µl - 35µl (0.1µl stepping)
Sampling Technology
Liquid level detection, clot detection and collision detection
Automatic Sample Dilution
Pre-dilution and automatic re-assay with diluted, reduced or increased sample volume. 3 - 170 times
Sample Probe
Dedicated sample probe with liquid level sensor, crash detection, bubble detection and clot detection. High pressure rinsing of the internal wall and
stream washing of external wall.
Sample Type
Serum, Plasma, Urine, CSF and Supernatant
Sample Addition
Sampling interruption for addition of samples
Reagent Unit
Reagent Probe
Dedicated twin reagent probes with liquid level sensor and crash detection. High pressure rinsing of the internal wall and stream washing of external
wall.
Reagent Volume
20µl - 350µl (0.1µl stepping)
Reagent Capacity
Double reagent carousel with 90 reagent positions suitable for 70ml and 20ml reagent bottles.
Reagent Cooling
Continuous cooling between 5-15oC
Reagent Identification
Automatic barcode reagent identification. Code 128
Reagent Inventory
Calculation of remaining reagent volume and number of tests available. Alerts for reagent shortage to ensure continuous analysis.
Reaction System
Reaction Cuvette
160 reusable plastic cuvettes, 5mm path length, 12 step cuvette washing system
Total Reaction Volume
120µl - 450µl
Reaction Temperature
37oC ±0.1oC
Reaction Time
10, 22 minutes
Reaction Disk Temperature
Circulating Water
Mixing
Dual flat paddle stirrers
Water Consumption
less than 30 litres per hour consumed at peak use
Wastewater Treatment
Two diffluence for high and low concentration wastewater. With the function of concentrated waste liquid level alarming
Optical System
Light Source
20W/12V halogen lamp with 2000 hours use
Photometer
Diffraction Grating Photometry
Detection Principle
12 wavelengths: 340nm, 380nm, 405nm, 450nm, 505nm, 546nm, 570nm, 600nm, 660nm, 700nm, 750nm and 800nm
Wavelength Precision
±2nm
OD Linear Range
0 - 3.3 Abs
Calibration and QC
Calibration
1 point linear method, 2 point linear method, multiple point linear method, non linear method, 9 types of calibration curve, Up to 6 calibrators per
test
Calibration Tracking
Automatically depicts K-value trends
Data Management
Real-time and monthly QC data archiving, Automatic plotting of Levy-Jennings charts, Alarms when control results out of range, Record reason for
discrepant results
Operating System
PC Operating System
Windows XP
Software
English version graphical operating software
Software Functions
Automatic calibration, Automatic barcode scanning, Reagent information management, Serum index, Reaction process monitoring, Dirty cup memory evading,
Cross contamination prevention, Patient information memory and association input, Automatic report audit, Report format statistic and printing,
Reference range classification, Alarming information classification, Automatic dormancy and start up, Real time online help
Report Printing
Report formats support the user-defined mode
PC configuration
CPU ≥2.2 GHz (dual-core processor); Memory ≥1G; Hard disk ≥ 160G; 17 inch LCD display; Stylus, inkjet, laser printer (optional)
System connection
TCP/IP network interface, standard RS-232C interface
Connectivity
Connect with LIS/HIS system
Power Requirements
Input Voltage
Voltage AC 220V ± 22V
Power Consumption
2500 VA, 50Hz ± 1Hz
Catalogue No./ Ordering Info
RX suzuka analyser RX8002
|